Mouse ELISA Kits

Mouse TRH(Thyrotropin Releasing Hormone) ELISA Kit
Copy Product Information

Product No

FY-EM2525

Sample Type

Serum, Plasma, Tissue homogenate and Other biological samples

Assay Time

1.5H

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
Mouse TRH(Thyrotropin Releasing Hormone) ELISA Kit
Species:
Mouse
Alternative Name:
Thyrotropin-releasing hormone, TRH, Protirelin, TSH-releasing factor, Thyroliberin, Thyrotropin-releasing factor, TRF
Uniport ID:
Q62361
Assay Type:
Competitive
Standard:
500pg/mL
Sensitivity:
4.69 pg/mL
Detection Range:
7.82-500pg/mL
Assay Time:
1.5H
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples
Research:
Tumor immunity;Endocrinology;Neuro science;Hepatology;Hormone metabolism;
Test Principle:
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with TRH protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to TRH. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of TRH in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
500 0.204 0.204 0.204
250 0.342 0.344 0.343
125 0.561 0.566 0.564
62.5 0.892 0.894 0.893
31.25 1.313 1.322 1.318
15.63 1.77 1.78 1.775
7.82 2.079 2.094 2.087
0 2.559 2.56 2.56
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 89-103 96
EDTA plasma (n=8) 80-93 85
Cell culture media (n=8) 82-95 88
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-102 81-94 82-96
Average (%) 101 93 91
1:4 Range (%) 86-97 81-94 80-92
Average (%) 88 84 81
1:8 Range (%) 89-97 96-105 83-96
Average (%) 95 96 94
1:16 Range (%) 80-95 92-103 95-107
Average (%) 95 95 101
Assay Procedures
Related literature
Related Products