Human ELISA Kits

Human IFN-γ(Interferon Gamma) ELISA Kit
Copy Product Information

Product No

FY-EH1402

Sample Type

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL

Assay Time

3.5H

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
Human IFN-γ(Interferon Gamma) ELISA Kit
Species:
Human
Alternative Name:
IFNG, IFG, IFI, Type II Interferon, IFNγ, INF-γ, INFγ, INFG, INF-G, INF-G
Uniport ID:
P01579
Assay Type:
Sandwich
Standard:
800pg/mL
Sensitivity:
2.94 pg/mL
Detection Range:
12.5-800pg/mL
Assay Time:
3.5H
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Test Principle:
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to IFN-γ. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to IFN-γ. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain IFN-γ, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of IFN-γ in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
800 2.444 2.465 2.455 2.418
400 1.593 1.604 1.599 1.567
200 0.951 0.955 0.953 0.925
100 0.443 0.446 0.445 0.417
50 0.216 0.218 0.217 0.19
25 0.106 0.106 0.106 0.08
12.5 0.061 0.061 0.061 0.035
0 0.026 0.026 0.026 0
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 88-106 97
EDTA plasma (n=8) 87-99 93
Cell culture media (n=8) 82-97 90
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 92-106 95-103 80-96
Average (%) 98 102 96
1:4 Range (%) 93-105 95-103 88-97
Average (%) 100 101 93
1:8 Range (%) 85-99 87-98 96-103
Average (%) 97 96 99
1:16 Range (%) 90-99 85-92 86-92
Average (%) 97 90 90
Assay Procedures
Related literature
Related Products