LTA4(Leukotriene A4) ELISA Kit

Product No

FY-EU1123

Sample Type

Serum, Plasma, Tissue homogenates, Cell culture supernates and Other biological fluids

Assay Time

1.5H

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
LTA4(Leukotriene A4) ELISA Kit
Species:
Universal
Alternative Name:
LT-A4
Assay Type:
Competitive
Standard:
10000 pg/mL
Sensitivity:
46.3 pg/mL
Detection Range:
156.25-10000 pg/mL
Assay Time:
1.5H
Sample Type:
Serum, Plasma, Tissue homogenates, Cell culture supernates and Other biological fluids
Research:
Metabolic pathway;Infection immunity;Autoimmunity;
Test Principle:
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with LTA4 protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to LTA4. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of LTA4 in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
10000 0.267 0.268 0.268
5000 0.367 0.369 0.368
2500 0.561 0.561 0.561
1250 0.822 0.829 0.826
625 1.117 1.118 1.118
312.5 1.512 1.522 1.517
156.25 1.826 1.836 1.831
0 2.49 2.508 2.499
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 82-95 88
EDTA plasma (n=8) 88-102 95
Cell culture media (n=8) 95-107 101
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 95-103 80-95 82-97
Average (%) 102 83 88
1:4 Range (%) 80-96 80-95 88-105
Average (%) 93 80 99
1:8 Range (%) 87-98 96-103 80-93
Average (%) 93 97 85
1:16 Range (%) 85-92 83-101 87-98
Average (%) 87 97 97