Universal ELISA Kits

VMA(Vanillylmandelic Acid) ELISA Kit

Product No

FY-EU1064

Sample Type

Urine and biological agents

Assay Time

1.5H

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
VMA(Vanillylmandelic Acid) ELISA Kit
Species:
Universal
Alternative Name:
Vanillyl Mandelic Acid
Assay Type:
Competitive
Standard:
10000 ng/mL
Sensitivity:
43.7 ng/mL
Detection Range:
156.25-10000 ng/mL
Assay Time:
1.5H
Sample Type:
Urine and biological agents
Research:
Metabolic pathway;Endocrinology;Cardiovascular biology;Hepatology;Gastroenterology;Hormone metabolism;
Test Principle:
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with VMA protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to VMA. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of VMA in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
10000 0.263 0.265 0.264
5000 0.363 0.366 0.365
2500 0.557 0.559 0.558
1250 0.818 0.823 0.821
625 1.113 1.122 1.118
312.5 1.508 1.512 1.51
156.25 1.822 1.839 1.831
0 2.486 2.509 2.498
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 84-96 90
EDTA plasma (n=8) 80-95 87
Cell culture media (n=8) 87-99 93
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 90-99 87-103 89-98
Average (%) 98 91 97
1:4 Range (%) 95-106 87-103 96-105
Average (%) 96 93 98
1:8 Range (%) 87-101 84-97 95-103
Average (%) 89 95 98
1:16 Range (%) 80-86 88-97 95-103
Average (%) 81 94 95
Assay Procedures
Related literature