EasyPro Series ELISA Kits

EasyPro Universal PGE2(Prostaglandin E2) ELISA Kit
Copy Product Information

Product No

FY-EU10131S

Sample Type

Serum, plasma and other biological fluids

Assay Time

45min

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
EasyPro Universal PGE2(Prostaglandin E2) ELISA Kit
Species:
Universal
Alternative Name:
PG-E2
Assay Type:
Competitive
Standard:
200pg/mL
Sensitivity:
0.97 pg/mL
Detection Range:
3.13-200 pg/mL
Assay Time:
45min
Sample Type:
Serum, plasma and other biological fluids
Research:
Metabolic pathway;Infection immunity;Endocrinology;Hormone metabolism;
Test Principle:
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with PGE2 Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for PGE2 are added to the micro ELISA plate wells. PGE2 in samples (or standards) competes with a fixed amount of PGE2 on the solid phase supporter for sites on the HRP linked detection antibody specific to PGE2. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of PGE2 in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
200 0.186 0.186 0.186
100 0.322 0.322 0.322
50 0.537 0.544 0.541
25 0.871 0.877 0.874
12.5 1.299 1.301 1.3
6.25 1.691 1.706 1.699
3.13 2.067 2.078 2.073
0 2.507 2.512 2.51
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-93 85
EDTA plasma (n=8) 87-99 93
Cell culture media (n=8) 90-103 96
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 83-94 82-94 91-99
Average (%) 86 82 96
1:4 Range (%) 85-96 82-94 80-92
Average (%) 85 89 80
1:8 Range (%) 92-103 85-96 86-102
Average (%) 101 94 89
1:16 Range (%) 86-97 80-90 87-103
Average (%) 93 88 91
Assay Procedures
Related literature
Related Products