EasyPro Series ELISA Kits

EasyPro Mouse β-CTx(Beta Crosslaps) ELISA Kit
Copy Product Information

Product No

FY-EM1540S

Sample Type

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL

Assay Time

45min

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
EasyPro Mouse β-CTx(Beta Crosslaps) ELISA Kit
Species:
Mouse
Alternative Name:
b-CTx, bCTXI, bCTX-I, B-Cr, BCL, Type I Collagen C-Telopeptide-Related Fraction
Assay Type:
Competitive
Standard:
5000pg/mL
Sensitivity:
46.88 pg/mL
Detection Range:
78.13-5000pg/mL
Assay Time:
45min
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Research:
Signal transduction;Hematology;
Test Principle:
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with β-CTx Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for β-CTx are added to the micro ELISA plate wells. β-CTx in samples (or standards) competes with a fixed amount of β-CTx on the solid phase supporter for sites on the HRP linked detection antibody specific to β-CTx. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of β-CTx in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
5000 0.162 0.163 0.163
2500 0.3 0.306 0.303
1250 0.519 0.52 0.52
625 0.85 0.862 0.856
312.5 1.271 1.288 1.28
156.25 1.728 1.738 1.733
78.13 2.037 2.072 2.055
0 2.517 2.528 2.523
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-92 85
EDTA plasma (n=8) 87-99 93
Cell culture media (n=8) 85-99 92
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-94 85-92 91-99
Average (%) 85 87 96
1:4 Range (%) 87-103 85-92 88-102
Average (%) 99 88 93
1:8 Range (%) 81-95 89-103 86-92
Average (%) 87 94 88
1:16 Range (%) 82-90 97-105 81-93
Average (%) 88 103 92
Assay Procedures
Related literature
Related Products