EasyPro Series ELISA Kits

EasyPro Rat Growth Regulated Oncogene Beta(GROβ/CXCL2)ELISA Kit

Product No

FY-ER3097S

Sample Type

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL

Assay Time

1.5H

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
EasyPro Rat Growth Regulated Oncogene Beta(GROβ/CXCL2)ELISA Kit
Species:
Rat
Alternative Name:
CXCL2, MIP-2a, GRO2, SCYB2, GROb, MGSA-B, CINC2a
Uniport ID:
P30348
Assay Type:
Sandwich
Standard:
2000pg/mL
Sensitivity:
18.75 pg/mL
Detection Range:
31.25-2000pg/mL
Assay Time:
1.5H
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL
Research:
Hematology;
Test Principle:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to GROβ/CXCL2. Samples (or Standards) and and Horseradish Peroxidase (HRP) linked antibody specific for GROβ/CXCL2 are added to the micro ELISA plate wells. GROβ/CXCL2 in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to GROβ/CXCL2. Excess conjugate and unbound sample or standard are washed from the plate. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of GROβ/CXCL2. You can calculate the concentration of GROβ/CXCL2 in the samples by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
2000 2.316 2.355 2.336 2.304
1000 1.522 1.544 1.533 1.501
500 0.891 0.904 0.898 0.866
250 0.408 0.409 0.409 0.377
125 0.229 0.23 0.23 0.198
62.5 0.129 0.13 0.13 0.098
31.25 0.082 0.083 0.083 0.051
0 0.032 0.032 0.032 0
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 87-99 93
EDTA plasma (n=8) 81-95 88
Cell culture media (n=8) 87-99 93
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-99 95-104 89-103
Average (%) 98 97 103
1:4 Range (%) 86-92 95-104 85-92
Average (%) 89 103 89
1:8 Range (%) 89-103 87-96 89-96
Average (%) 94 92 89
1:16 Range (%) 97-105 82-96 90-98
Average (%) 98 92 94
Assay Procedures
Related literature
Related Products