Product Details
Product Name in English:
EasyPro Human S100B(S100 Calcium Binding Protein B) ELISA Kit
Alternative Name:
S1-B, NEF, S1Beta
Detection Range:
31.25-2000pg/mL
Sample Type:
cerebrospinal fluid, blood, Urine, amniotic fluid;Sample Volume=50uL
Research:
Signal transduction;
Test Principle:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to S100B. Samples (or Standards) and and Horseradish Peroxidase (HRP) linked antibody specific for S100B are added to the micro ELISA plate wells. S100B in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to S100B. Excess conjugate and unbound sample or standard are washed from the plate. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of S100B. You can calculate the concentration of S100B in the samples by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration |
OD1 |
OD2 |
Average OD |
Corrected |
2000 |
2.349 |
2.372 |
2.361 |
2.295 |
1000 |
1.555 |
1.556 |
1.556 |
1.49 |
500 |
0.924 |
0.942 |
0.933 |
0.867 |
250 |
0.441 |
0.447 |
0.444 |
0.378 |
125 |
0.262 |
0.265 |
0.264 |
0.198 |
62.5 |
0.162 |
0.165 |
0.164 |
0.098 |
31.25 |
0.115 |
0.115 |
0.115 |
0.049 |
0 |
0.065 |
0.066 |
0.066 |
0 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item |
Intra-assay Precision |
Intra-assay Precision |
Sample |
3 |
3 |
Replicate |
9 |
18 |
CV(%) |
5 |
8 |
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type |
Range(%) |
Average Recovery(%) |
Serum (n=8) |
84-93 |
89 |
EDTA plasma (n=8) |
81-103 |
87 |
Cell culture media (n=8) |
93-105 |
99 |
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
|
|
Serum (n=5) |
EDTA plasma (n=5) |
Cell culture media (n=5) |
1:2 |
Range (%) |
88-104 |
97-105 |
80-97 |
Average (%) |
96 |
99 |
95 |
1:4 |
Range (%) |
93-102 |
84-96 |
90-99 |
Average (%) |
101 |
87 |
98 |
1:8 |
Range (%) |
82-97 |
87-98 |
95-103 |
Average (%) |
95 |
94 |
100 |
1:16 |
Range (%) |
92-101 |
95-102 |
89-97 |
Average (%) |
95 |
99 |
93 |