EasyPro Series ELISA Kits

EasyPro Human Macrophage Inflammatory Protein 3β(MIP-3β/ELC/CCL19)ELISA Kit

Product No

FY-EH2800S

Sample Type

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL

Assay Time

1.5H

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
EasyPro Human Macrophage Inflammatory Protein 3β(MIP-3β/ELC/CCL19)ELISA Kit
Species:
Human
Alternative Name:
CCL19, CKb11, ELC, MIP-3b, MIP3B, SCYA19, C-C motif chemokine ligand 19
Uniport ID:
Q99731
Assay Type:
Sandwich
Standard:
500pg/mL
Sensitivity:
4.69 pg/mL
Detection Range:
7.82-500pg/mL
Assay Time:
1.5H
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL
Research:
Enzyme & Kinase;
Test Principle:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to MIP-3β/ELC/CCL19. Samples (or Standards) and and Horseradish Peroxidase (HRP) linked antibody specific for MIP-3β/ELC/CCL19 are added to the micro ELISA plate wells. MIP-3β/ELC/CCL19 in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to MIP-3β/ELC/CCL19. Excess conjugate and unbound sample or standard are washed from the plate. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of MIP-3β/ELC/CCL19. You can calculate the concentration of MIP-3β/ELC/CCL19 in the samples by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
500 2.471 2.514 2.493 2.457
250 1.615 1.622 1.619 1.583
125 0.949 0.952 0.951 0.915
62.5 0.43 0.433 0.432 0.396
31.25 0.221 0.221 0.221 0.185
15.63 0.123 0.124 0.124 0.088
7.82 0.071 0.072 0.072 0.036
0 0.036 0.036 0.036 0
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 89-101 95
EDTA plasma (n=8) 82-95 88
Cell culture media (n=8) 80-94 87
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 87-98 93-102 82-95
Average (%) 92 96 86
1:4 Range (%) 92-101 93-102 88-103
Average (%) 96 94 92
1:8 Range (%) 82-98 81-96 86-97
Average (%) 92 82 93
1:16 Range (%) 89-97 93-106 87-102
Average (%) 94 97 92
Assay Procedures
Related literature
Related Products