EasyPro Series ELISA Kits

EasyPro Universal Androstenedione(ASD)ELISA Kit

Product No

FY-EU10137S

Sample Type

serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids

Assay Time

45min

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
EasyPro Universal Androstenedione(ASD)ELISA Kit
Species:
Universal
Alternative Name:
4-Androstenedione; 4-Androstene-3,17-Dione
Assay Type:
Competitive
Standard:
10000pg/mL
Sensitivity:
48.8 pg/mL
Detection Range:
156.25-10000 pg/mL
Assay Time:
45min
Sample Type:
serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Research:
Endocrinology;Reproductive science;Hormone metabolism;
Test Principle:
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with ASD Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for ASD are added to the micro ELISA plate wells. ASD in samples (or standards) competes with a fixed amount of ASD on the solid phase supporter for sites on the HRP linked detection antibody specific to ASD. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of ASD in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
10000 0.279 0.284 0.282
5000 0.379 0.383 0.381
2500 0.573 0.575 0.574
1250 0.834 0.841 0.838
625 1.129 1.133 1.131
312.5 1.524 1.545 1.535
156.25 1.838 1.865 1.852
0 2.502 2.518 2.51
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 84-103 96
EDTA plasma (n=8) 82-96 87
Cell culture media (n=8) 86-97 92
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 83-97 95-103 95-104
Average (%) 89 99 100
1:4 Range (%) 84-99 80-96 82-93
Average (%) 96 87 91
1:8 Range (%) 95-102 93-106 81-96
Average (%) 100 103 93
1:16 Range (%) 88-103 87-98 83-94
Average (%) 88 98 92
Assay Procedures
Related literature
Related Products