EasyPro Series ELISA Kits

EasyPro Human Gastrin(GT)ELISA Kit

Product No

FY-EH2209S

Sample Type

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL

Assay Time

45min

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
EasyPro Human Gastrin(GT)ELISA Kit
Species:
Human
Alternative Name:
GAST, GAS, gastrin
Uniport ID:
P01350
Assay Type:
Competitive
Standard:
1000pg/mL
Sensitivity:
9.38 pg/mL
Detection Range:
15.63-1000pg/mL
Assay Time:
45min
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Research:
Signal transduction;Infection immunity;Neuro science;
Test Principle:
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with GT Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for GT are added to the micro ELISA plate wells. GT in samples (or standards) competes with a fixed amount of GT on the solid phase supporter for sites on the HRP linked detection antibody specific to GT. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of GT in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
1000 0.303 0.304 0.304
500 0.403 0.41 0.407
250 0.597 0.597 0.597
125 0.858 0.87 0.864
62.5 1.153 1.166 1.16
31.25 1.548 1.554 1.551
15.63 1.862 1.877 1.87
0 2.526 2.572 2.549
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 81-95 88
EDTA plasma (n=8) 88-96 92
Cell culture media (n=8) 95-107 101
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 93-101 92-105 88-92
Average (%) 95 95 89
1:4 Range (%) 82-94 92-105 86-97
Average (%) 91 96 96
1:8 Range (%) 86-96 83-92 90-101
Average (%) 91 89 93
1:16 Range (%) 81-97 86-93 82-96
Average (%) 87 88 87
Assay Procedures
Related literature
Related Products