High Sensitive ELISA Kits

High Sensitivity Human EGF(Epidermal Growth Factor) ELISA Kit
Copy Product Information

Product No

FY-HS-EH1270

Sample Type

3.5H

Assay Time

0.16-10pg/mL

Price($)

Size

Purchase details

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

Product Details
Product Name in English:
High Sensitivity Human EGF(Epidermal Growth Factor) ELISA Kit
Species:
FY-HS-EH1270-3
Alternative Name:
Human
Uniport ID:
URG, HOMG4, Beta-Urogastrone
Assay Type:
P01133
Standard:
Sandwich
Sensitivity:
10pg/mL
Detection Range:
0.05 pg/mL
Assay Time:
0.16-10pg/mL
Sample Type:
3.5H
Research:
Serum, Plasma and Other biological fluids;Sample Volume=100μL
Test Principle:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human EGF. Samples (or Standards) and and Horseradish Peroxidase (HRP) linked antibody specific for Human EGF are added to the micro ELISA plate wells. Human EGF in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human EGF. Excess conjugate and unbound sample or standard are washed from the plate. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of Human EGF. You can calculate the concentration of Human EGF in the samples by comparing the OD of the samples to the standard curve.
Technical Data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
10 2.282 2.472 2.377 2.358
5 1.675 1.709 1.692 1.673
2.5 0.956 0.996 0.976 0.957
1.25 0.46 0.478 0.469 0.45
0.63 0.239 0.221 0.23 0.211
0.32 0.105 0.109 0.107 0.088
0.16 0.059 0.061 0.06 0.041
0 0.019 0.019 0.019 0
Linear
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 86-98 89-114 92-109
Average (%) 93 109 93
1:4 Range (%) 90-111 91-106 90-114
Average (%) 92 105 111
1:8 Range (%) 88-111 92-113 91-108
Average (%) 102 108 100
1:16 Range (%) 89-113 91-110 92-99
Average (%) 101 109 97
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 91-100 99
EDTA plasma (n=8) 93-112 106
Cell culture media (n=8) 89-105 102