人ELISA试剂盒

人心钠肽(ANP)酶联免疫吸附测定试剂盒
复制

货号

FY-EH1890

样本类型

Plasma

检测时间

1.5H

价格(美元)

规格

采购详情

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

产品详情
英文名称:
Human ANP(Atrial Natriuretic Peptide) ELISA Kit
种属:
Human
别称:
NPPA, ANF, ANP, ATFB6, ATRST2, CDD, CDD-ANF, CDP, PND, Atrial natriuretic peptide, natriuretic peptide A
Uniport ID:
P01160(124~151aa)
检测方法:
Competitive
标准品:
500pg/mL
灵敏度:
4.69 pg/mL
检测范围:
7.82-500pg/mL
反应时间:
1.5H
样品类型:
Plasma
研究领域:
Tumor immunity;
测试原理:
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with ANP protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to ANP. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of ANP in the samples is then determined by comparing the OD of the samples to the standard curve.
技术数据
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
500 0.18 0.18 0.18
250 0.318 0.321 0.32
125 0.537 0.539 0.538
62.5 0.868 0.87 0.869
31.25 1.289 1.3 1.295
15.63 1.746 1.748 1.747
7.82 2.055 2.064 2.06
0 2.535 2.551 2.543
精密度
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
回收率
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
样本类型 Range(%) Average Recovery(%)
Serum (n=8) 88-101 94
EDTA plasma (n=8) 80-91 84
Cell culture media (n=8) 80-96 88
线性
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 91-99 85-100 88-102
Average (%) 93 90 92
1:4 Range (%) 83-97 85-100 86-93
Average (%) 85 89 88
1:8 Range (%) 87-105 84-99 93-101
Average (%) 91 92 97
1:16 Range (%) 87-99 85-97 85-98
Average (%) 91 93 94
检测流程图
相关文献
相关产品