产品详情
英文名称:
E1(Estrone) ELISA Kit
样品类型:
Serum, Plasma and Other biological fluids
研究领域:
Metabolic pathway;Endocrinology;Reproductive science;Genetic science;Hormone metabolism;
测试原理:
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with E1 protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to E1. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of E1 in the samples is then determined by comparing the OD of the samples to the standard curve.
技术数据
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration |
OD1 |
OD2 |
Average OD |
Corrected |
2000 |
0.182 |
0.183 |
0.183 |
|
1000 |
0.318 |
0.319 |
0.319 |
|
500 |
0.533 |
0.537 |
0.535 |
|
250 |
0.867 |
0.868 |
0.868 |
|
125 |
1.295 |
1.295 |
1.295 |
|
62.5 |
1.687 |
1.696 |
1.692 |
|
31.25 |
2.063 |
2.08 |
2.072 |
|
0 |
2.503 |
2.517 |
2.51 |
|
精密度
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item |
Intra-assay Precision |
Intra-assay Precision |
Sample |
3 |
3 |
Replicate |
9 |
18 |
CV(%) |
5 |
8 |
回收率
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
样本类型 |
Range(%) |
Average Recovery(%) |
Serum (n=8) |
86-99 |
92 |
EDTA plasma (n=8) |
82-95 |
88 |
Cell culture media (n=8) |
80-95 |
87 |
线性
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
|
|
Serum (n=5) |
EDTA plasma (n=5) |
Cell culture media (n=5) |
1:2 |
Range (%) |
85-105 |
86-93 |
90-101 |
Average (%) |
90 |
92 |
95 |
1:4 |
Range (%) |
87-98 |
86-93 |
80-101 |
Average (%) |
88 |
89 |
87 |
1:8 |
Range (%) |
91-102 |
93-101 |
87-96 |
Average (%) |
92 |
94 |
90 |
1:16 |
Range (%) |
86-102 |
86-97 |
96-105 |
Average (%) |
89 |
88 |
103 |