通用型ELISA试剂盒

维生素D3(VD3)酶联免疫吸附测定试剂盒
复制

货号

FY-EU1114

样本类型

Serum, Plasma and Other biological fluids

检测时间

1.5H

价格(美元)

规格

采购详情

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

产品详情
英文名称:
VD3(Vitamin D3) ELISA Kit
种属:
Universal
别称:
Cholecalciferol; Toxiferol; Activated 7-Dehydrocholesterol
检测方法:
Competitive
标准品:
400 ng/mL
灵敏度:
2.14 ng/mL
检测范围:
6.25-400 ng/mL
反应时间:
1.5H
样品类型:
Serum, Plasma and Other biological fluids
研究领域:
Metabolic pathway;Reproductive science;Nutrition metabolism;Bone metabolism;
测试原理:
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with VD3 protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to VD3. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of VD3 in the samples is then determined by comparing the OD of the samples to the standard curve.
技术数据
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
400 0.258 0.258 0.258
200 0.421 0.421 0.421
100 0.661 0.664 0.663
50 0.978 0.988 0.983
25 1.271 1.275 1.273
12.5 1.591 1.592 1.592
6.25 1.861 1.863 1.862
0 2.683 2.704 2.694
精密度
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
回收率
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
样本类型 Range(%) Average Recovery(%)
Serum (n=8) 92-105 98
EDTA plasma (n=8) 84-96 90
Cell culture media (n=8) 80-95 87
线性
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 86-93 93-102 91-104
Average (%) 91 98 92
1:4 Range (%) 91-105 93-102 89-97
Average (%) 96 101 93
1:8 Range (%) 87-98 93-106 93-102
Average (%) 90 102 97
1:16 Range (%) 89-94 87-101 81-96
Average (%) 94 90 88
检测流程图
相关文献
相关产品