产品详情
英文名称:
EasyPro Human HP(Haptoglobin) ELISA Kit
别称:
Hpt, BP, Hp2-Alpha, HPA1S
样品类型:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL
研究领域:
CD & Adhesion molecule;
测试原理:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to HP. Samples (or Standards) and and Horseradish Peroxidase (HRP) linked antibody specific for HP are added to the micro ELISA plate wells. HP in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to HP. Excess conjugate and unbound sample or standard are washed from the plate. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of HP. You can calculate the concentration of HP in the samples by comparing the OD of the samples to the standard curve.
技术数据
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration |
OD1 |
OD2 |
Average OD |
Corrected |
500 |
2.457 |
2.5 |
2.479 |
2.457 |
250 |
1.601 |
1.612 |
1.607 |
1.585 |
125 |
0.935 |
0.948 |
0.942 |
0.92 |
62.5 |
0.416 |
0.419 |
0.418 |
0.396 |
31.25 |
0.207 |
0.209 |
0.208 |
0.186 |
15.63 |
0.109 |
0.11 |
0.11 |
0.088 |
7.82 |
0.057 |
0.057 |
0.057 |
0.035 |
0 |
0.022 |
0.022 |
0.022 |
0 |
精密度
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item |
Intra-assay Precision |
Intra-assay Precision |
Sample |
3 |
3 |
Replicate |
9 |
18 |
CV(%) |
5 |
8 |
回收率
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
样本类型 |
Range(%) |
Average Recovery(%) |
Serum (n=8) |
92-105 |
98 |
EDTA plasma (n=8) |
84-96 |
90 |
Cell culture media (n=8) |
80-95 |
87 |
线性
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
|
|
Serum (n=5) |
EDTA plasma (n=5) |
Cell culture media (n=5) |
1:2 |
Range (%) |
86-93 |
93-102 |
91-104 |
Average (%) |
91 |
98 |
92 |
1:4 |
Range (%) |
91-105 |
93-102 |
89-97 |
Average (%) |
96 |
101 |
93 |
1:8 |
Range (%) |
87-98 |
93-106 |
93-102 |
Average (%) |
90 |
102 |
97 |
1:16 |
Range (%) |
89-94 |
87-101 |
81-96 |
Average (%) |
94 |
90 |
88 |