快速检测系列ELISA试剂盒

一步法人粒细胞巨噬细胞集落刺激因子(GM-CSF)酶联免疫吸附测定试剂盒

货号

FY-EH2862S

样本类型

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL

检测时间

1.5H

价格(美元)

规格

采购详情

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

产品详情
英文名称:
EasyPro Human Granulocyte-Macrophage Colony Stimulating Factor(GM-CSF)ELISA Kit
种属:
Human
别称:
GMCSF, CSF2
Uniport ID:
P04141
检测方法:
Sandwich
标准品:
500pg/mL
灵敏度:
4.69 pg/mL
检测范围:
7.82-500pg/mL
反应时间:
1.5H
样品类型:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL
研究领域:
Tumor immunity;
测试原理:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to GM-CSF. Samples (or Standards) and and Horseradish Peroxidase (HRP) linked antibody specific for GM-CSF are added to the micro ELISA plate wells. GM-CSF in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to GM-CSF. Excess conjugate and unbound sample or standard are washed from the plate. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of GM-CSF. You can calculate the concentration of GM-CSF in the samples by comparing the OD of the samples to the standard curve.
技术数据
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
500 2.468 2.491 2.48 2.446
250 1.612 1.633 1.623 1.589
125 0.946 0.958 0.952 0.918
62.5 0.427 0.433 0.43 0.396
31.25 0.218 0.219 0.219 0.185
15.63 0.12 0.121 0.121 0.087
7.82 0.068 0.068 0.068 0.034
0 0.033 0.034 0.034 0
精密度
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
回收率
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
样本类型 Range(%) Average Recovery(%)
Serum (n=8) 80-90 84
EDTA plasma (n=8) 85-97 91
Cell culture media (n=8) 80-93 86
线性
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-96 86-101 92-105
Average (%) 87 101 104
1:4 Range (%) 87-101 86-101 89-102
Average (%) 97 100 90
1:8 Range (%) 89-102 86-101 92-101
Average (%) 90 92 95
1:16 Range (%) 85-98 87-96 88-102
Average (%) 91 89 94
检测流程图
相关文献
相关产品