快速检测系列ELISA试剂盒

一步法人肠脂肪酸结合蛋白(IFABP/FABP2)酶联免疫吸附测定试剂盒
复制

货号

FY-EH2870S

样本类型

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL

检测时间

1.5H

价格(美元)

规格

采购详情

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

产品详情
英文名称:
EasyPro Human IFABP/FABP2(Intestinal Fatty Acid Binding Protein) ELISA Kit
种属:
Human
别称:
FABPI, I-FABP
Uniport ID:
P12104
检测方法:
Sandwich
标准品:
10ng/mL
灵敏度:
0.1 ng/mL
检测范围:
0.16-10ng/mL
反应时间:
1.5H
样品类型:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL
研究领域:
Enzyme & Kinase;
测试原理:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to IFABP/FABP2. Samples (or Standards) and and Horseradish Peroxidase (HRP) linked antibody specific for IFABP/FABP2 are added to the micro ELISA plate wells. IFABP/FABP2 in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to IFABP/FABP2. Excess conjugate and unbound sample or standard are washed from the plate. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of IFABP/FABP2. You can calculate the concentration of IFABP/FABP2 in the samples by comparing the OD of the samples to the standard curve.
技术数据
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
10 2.417 2.452 2.435 2.375
5 1.732 1.756 1.744 1.684
2.5 1.016 1.034 1.025 0.965
1.25 0.509 0.51 0.51 0.45
0.63 0.27 0.275 0.273 0.213
0.32 0.147 0.149 0.148 0.088
0.16 0.1 0.102 0.101 0.041
0 0.059 0.06 0.06 0
精密度
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
回收率
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
样本类型 Range(%) Average Recovery(%)
Serum (n=8) 92-105 98
EDTA plasma (n=8) 83-95 89
Cell culture media (n=8) 80-93 86
线性
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 87-93 84-99 86-105
Average (%) 90 91 105
1:4 Range (%) 91-103 81-95 84-98
Average (%) 94 90 95
1:8 Range (%) 82-99 98-107 95-102
Average (%) 97 103 96
1:16 Range (%) 99-104 96-101 88-106
Average (%) 103 101 91
检测流程图