快速检测系列ELISA试剂盒

一步法人免疫球蛋白G(IgG)酶联免疫吸附测定试剂盒

货号

FY-EH2873S

样本类型

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL

检测时间

1.5H

价格(美元)

规格

采购详情

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

产品详情
英文名称:
EasyPro Human Immunoglobulin G(IgG)ELISA Kit
种属:
Human
别称:
IgG
Uniport ID:
P01860(IGHG3)
检测方法:
Sandwich
标准品:
100ng/mL
灵敏度:
0.94 ng/mL
检测范围:
1.57-100ng/mL
反应时间:
1.5H
样品类型:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL
研究领域:
Signal transduction;
测试原理:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to IgG. Samples (or Standards) and and Horseradish Peroxidase (HRP) linked antibody specific for IgG are added to the micro ELISA plate wells. IgG in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to IgG. Excess conjugate and unbound sample or standard are washed from the plate. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of IgG. You can calculate the concentration of IgG in the samples by comparing the OD of the samples to the standard curve.
技术数据
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
100 2.346 2.388 2.367 2.332
50 1.544 1.558 1.551 1.516
25 0.941 0.945 0.943 0.908
12.5 0.509 0.518 0.514 0.479
6.25 0.254 0.257 0.256 0.221
3.13 0.139 0.14 0.14 0.105
1.57 0.078 0.079 0.079 0.044
0 0.035 0.035 0.035 0
精密度
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
回收率
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
样本类型 Range(%) Average Recovery(%)
Serum (n=8) 92-106 99
EDTA plasma (n=8) 80-95 86
Cell culture media (n=8) 88-103 95
线性
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 87-96 80-93 87-101
Average (%) 88 87 93
1:4 Range (%) 92-105 80-93 86-94
Average (%) 94 87 86
1:8 Range (%) 89-102 89-97 85-95
Average (%) 94 92 93
1:16 Range (%) 92-101 85-94 89-99
Average (%) 98 85 90
检测流程图
相关文献
相关产品