大鼠ELISA试剂盒

大鼠皮肤T细胞虏获趋化因子(CTACK)酶联免疫吸附测定试剂盒
复制

货号

FY-ER1218

样本类型

Serum, Plasma, Tissue homogenate and Other biological samples

检测时间

3.5H

价格(美元)

规格

采购详情

Contact supplier to negotiate price details

100% Quality Assurance

Fast Shipping

产品详情
英文名称:
Rat CTACK(Cutaneous T-Cell Attracting Chemokine) ELISA Kit
种属:
Rat
别称:
CCL27, SCYA27, ALP, ILC, Skinkine, ESkine, PESKY, CTAK
检测方法:
Sandwich
标准品:
4000pg/mL
灵敏度:
37.5 pg/mL
检测范围:
62.5-4000pg/mL
反应时间:
3.5H
样品类型:
Serum, Plasma, Tissue homogenate and Other biological samples
研究领域:
Signal transduction;Tumor immunity;
测试原理:
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to CTACK. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to CTACK. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain CTACK, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of CTACK in the samples is then determined by comparing the OD of the samples to the standard curve.
技术数据
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration OD1 OD2 Average OD Corrected
4000 2.399 2.421 2.41 2.334
2000 1.559 1.562 1.561 1.494
1000 0.885 0.886 0.886 0.82
500 0.485 0.486 0.486 0.42
250 0.265 0.265 0.265 0.2
125 0.162 0.162 0.162 0.097
62.5 0.115 0.115 0.115 0.05
0 0.065 0.065 0.065 0
精密度
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item Intra-assay Precision Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8
回收率
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
样本类型 Range(%) Average Recovery(%)
Serum (n=8) 80-90 85
EDTA plasma (n=8) 93-106 100
Cell culture media (n=8) 81-93 87
线性
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 87-96 87-98 88-99
Average (%) 89 94 98
1:4 Range (%) 82-95 87-100 81-93
Average (%) 91 94 93
1:8 Range (%) 93-102 85-94 89-97
Average (%) 99 90 92
1:16 Range (%) 87-96 83-96 85-94
Average (%) 89 85 91
检测流程图
相关文献
相关产品